ACUTE PULMONARY EMBOLISM TREATED WITH TISSUE PLASMINOGEN ACTIVATOR

The Lancet ◽  
1986 ◽  
Vol 328 (8512) ◽  
pp. 886-889 ◽  
Author(s):  
SamuelZ. Goldhaber ◽  
JohnE. Markis ◽  
MichaelF. Meyerovitz ◽  
DuckSoo Kim ◽  
DouglasL. Dawley ◽  
...  
VASA ◽  
2014 ◽  
Vol 43 (6) ◽  
pp. 450-458 ◽  
Author(s):  
Julio Flores ◽  
Ángel García-Avello ◽  
Esther Alonso ◽  
Antonio Ruíz ◽  
Olga Navarrete ◽  
...  

Background: We evaluated the diagnostic efficacy of tissue plasminogen activator (tPA), using an enzyme-linked immunosorbent assay (ELISA) and compared it with an ELISA D-dimer (VIDAS D-dimer) in acute pulmonary embolism (PE). Patients and methods: We studied 127 consecutive outpatients with clinically suspected PE. The diagnosis of PE was based on a clinical probability pretest for PE and a strict protocol of imaging studies. A plasma sample to measure the levels of tPA and D-dimer was obtained at enrollment. Diagnostic accuracy for tPA and D-dimer was determined by the area under the receiver operating characteristic (ROC) curve. Sensitivity, specificity, predictive values, and the diagnostic utility of tPA with a cutoff of 8.5 ng/mL and D-dimer with a cutoff of 500 ng/mL, were calculated for PE diagnosis. Results: PE was confirmed in 41 patients (32 %). Areas under ROC curves were 0.86 for D-dimer and 0.71 for tPA. The sensitivity/negative predictive value for D-dimer using a cutoff of 500 ng/mL, and tPA using a cutoff of 8.5 ng/mL, were 95 % (95 % CI, 88–100 %)/95 % (95 % CI, 88–100 %) and 95 % (95 % CI, 88–100 %)/94 %), respectively. The diagnostic utility to exclude PE was 28.3 % (95 % CI, 21–37 %) for D-dimer and 24.4 % (95 % CI, 17–33 %) for tPA. Conclusions: The tPA with a cutoff of 8.5 ng/mL has a high sensitivity and negative predictive value for exclusion of PE, similar to those observed for the VIDAS D-dimer with a cutoff of 500 ng/mL, although the diagnostic utility was slightly higher for the D-dimer.


Sign in / Sign up

Export Citation Format

Share Document