scholarly journals Performance of EasyBreath Decathlon Snorkeling mask for delivering continuous positive airway pressure

2021 ◽  
Vol 11 (1) ◽  
Author(s):  
Alberto Noto ◽  
Claudia Crimi ◽  
Andrea Cortegiani ◽  
Massimiliano Giardina ◽  
Filippo Benedetto ◽  
...  

AbstractDuring the COVID-19 pandemic, the need for noninvasive respiratory support devices has dramatically increased, sometimes exceeding hospital capacity. The full-face Decathlon snorkeling mask, EasyBreath (EB mask), has been adapted to deliver continuous positive airway pressure (CPAP) as an emergency respiratory interface. We aimed to assess the performance of this modified EB mask and to test its use during different gas mixture supplies. CPAP set at 5, 10, and 15 cmH2O was delivered to 10 healthy volunteers with a high-flow system generator set at 40, 80, and 120 L min−1 and with a turbine-driven ventilator during both spontaneous and loaded (resistor) breathing. Inspiratory CO2 partial pressure (PiCO2), pressure inside the mask, breathing pattern and electrical activity of the diaphragm (EAdi) were measured at all combinations of CPAP/flows delivered, with and without the resistor. Using the high-flow generator set at 40 L min−1, the PiCO2 significantly increased and the system was unable to maintain the target CPAP of 10 and 15 cmH2O and a stable pressure within the respiratory cycle; conversely, the turbine-driven ventilator did. EAdi significantly increased with flow rates of 40 and 80 L min−1 but not at 120 L min−1 and with the turbine-driven ventilator. EB mask can be safely used to deliver CPAP only under strict constraints, using either a high-flow generator at a flow rate greater than 80 L min−1, or a high-performance turbine-driven ventilator.

2020 ◽  
Author(s):  
Alberto Noto ◽  
Claudia Crimi ◽  
Andrea Cortegiani ◽  
Massimiliano Giardina ◽  
Filippo Benedetto ◽  
...  

ABSTRACTBackgroundDuring the COVID-19 pandemic, the need for noninvasive respiratory support devices has dramatically increased, sometimes exceeding hospital capacity. The full-face Decathlon snorkeling mask, EasyBreath® (EB® mask), has been adapted to deliver continuous positive airway pressure (CPAP) as an emergency respiratory interface. We aimed to assess the performance of this modified EB® mask.MethodsCPAP set at 5, 10, and 15 cmH2O was delivered to 10 healthy volunteers with a high-flow system generator set at 40, 80, and 120 L min-1 and with a turbine-driven ventilator during both spontaneous and loaded (resistor) breathing. Inspiratory CO2 partial pressure (PiCO2), pressure inside the mask, breathing pattern and electrical activity of the diaphragm (EAdi) were measured at all combinations of CPAP/flows delivered, with and without the resistor.ResultsUsing the high-flow generator set at 40 L min-1, the PiCO2 significantly increased and the system was unable to maintain the target CPAP of 10 and 15 cmH2O and a stable pressure within the respiratory cycle; conversely, the turbine-driven ventilator did. EAdi significantly increased with flow rates of 40 and 80 L min-1 but not at 120 L min-1 and with the turbine-driven ventilator.ConclusionsEB® mask can be safely used to deliver CPAP only under strict constraints, using either a high-flow generator at a flow rate greater than 80 L min-1, or a high-performance turbine-driven ventilator.


Pneumologie ◽  
2021 ◽  
Author(s):  
Wolfram Windisch ◽  
Bernd Schönhofer ◽  
Daniel Sebastian Majorski ◽  
Maximilian Wollsching-Strobel ◽  
Carl-Peter Criée ◽  
...  

ZusammenfassungIn der Corona-Pandemie werden zunehmend nicht-invasive Verfahren zur Behandlung des akuten hypoxämischen Versagens bei COVID-19 eingesetzt. Hier stehen mit der HFOT (high-flow oxygen therapy), CPAP (continuous positive airway pressure) und der NIV (non-invasive ventilation) unterschiedliche Verfahren zur Verfügung, die das Ziel einer Intubationsvermeidung verfolgen. Der aktuelle Übersichtsartikel fasst die heterogene Studienlage zusammen. Wesentlich ist die Erkenntnis, dass diese nicht-invasiven Verfahren durchaus auch bei einem schweren, akuten hypoxämischen Versagen erfolgreich sein können und damit die Intubation wie auch Tubus-assoziierte Komplikationen vermeiden können. Demgegenüber bleibt aber ebenso zu betonen, dass die prolongierte unterstützte Spontanatmung ebenfalls zu Komplikationen führt und dass demzufolge insbesondere ein spätes NIV-Versagen mit erheblich verschlechterter Prognose einhergeht, was vor dem Hintergrund weiterhin hoher NIV-Versagensraten in Deutschland bedeutsam ist. Der aktuelle Artikel verweist schließlich auch auf einen Parallelartikel in dieser Ausgabe, der die medial in der Öffentlichkeit in Deutschland geführte Debatte zu diesem Thema aufgreift und deren inhaltliche Fragwürdigkeit, aber auch die negativen Auswirkungen auf die Gesellschaft und die Fachwelt adressiert. Gleichzeitig wird die Bedeutung von regelmäßig zu überarbeitenden Leitlinien untermauert.


Sign in / Sign up

Export Citation Format

Share Document