THE USE OF NITINOL RODS IN SURGICAL TREATMENT OF DEGENERATIVE SCOLIOSIS. 2.5-YEAR FOLLOW-UP
ABSTRACT Objectives: To compare the outcomes of surgical treatment with lumbar fixation using nitinol rods without fusion and with standard lumbar fixation with titanium rods and interbody fusion. Methods: Treatment results of 70 patients with degenerative lumbar scoliosis aged 40 to 82 were analyzed. In all cases pedicle screws and nitinol rods with a diameter of 5.5 mm were used. Thirty patients underwent fixation at L1-S1 and 40 patients underwent fixation at L1-L5. Spinal fusion was not performed. All patients had radiography, CT and MRI performed. The results were assessed according to the Oswestry scale, SRS 22, SF 36 and VAS. The minimum follow-up period for all patients was 2.5 years. For the control group, consisting of 72 patients, pedicle fixation with titanium rods and interbody fusion in the lumbosacral region were performed. Results: The average level of deformity correction equaled 25° (10° - 38°). The analysis of X-ray and CT-scans revealed a single patient with implant instability, two patients with bone resorption around the screws and one patient with rod fractures. Functional radiography 2.5 years after surgery showed an average mobility of the lumbar spine of 21° (15° - 30°). There were no problems at the adjacent levels. Conclusions: The use of nitinol rods in spinal deformity surgery is promising. This technology is an alternative to rigid fixation. Continued gathering of clinical data and its further evaluation is necessary.