P1793Eight-year clinical outcome and valve durability after transcatheter aortic valve implantation
Abstract Background With the growing indications of transcatheter aortic valve implantation (TAVI) worldwide and among lower risk patients, valve durability has become a crucial issue. Purpose To assess mid and long-term evolution of different generations of percutaneous balloon-expandable prostheses, predictive factors of valve deterioration and its correlation with long-term mortality. Methods All consecutive patients undergoing TAVI for severe aortic stenosis with balloon-expandable prosthesis between 2009 and 2014 and with a minimum follow-up of one-year were included in this monocentric prospective study. All echocardiograms were reviewed by two independent experts. Clinical events were defined according to the Valve Academic Research Consortium criteria. Valve deterioration was defined according to the 2017 EAPCI-ESC-EACTS international consensus statement at the longest follow-up. Results A total of 160 patients were included with a median follow-up of 3.4 years [1.5–4.9] and a maximum of 8 years. Patients were mostly implanted with the first generation Sapien XT valve (n=138, 86.2%). Median age was 85 [79–86] years, with 42.5% of women and a median logistic Euro-SCORE of 14.2% [10.6–23.2]. Immediately after TAVI, mean aortic gradient decreased dramatically from 51±12mmHg to 9±2.6mmHg (p<0.0001) and remained overall stable with a mean gradient of 12±1mmHg at 8 years. Valve deterioration occurred in 5.6% (n=9) of patients, of which 3.7% (n=6) with severe deterioration. Moderate or severe peri-prosthetic aortic regurgitation was observed in 2.5% (n=4) of patients. The eight-year survival rate was 12.9%. During follow-up, hospitalization for acute heart failure was required for 23.7% (n=38) of patients, a myocardial infarction or a stroke occurred respectively among 1.9% (n=3) and 5% (n=8) of patients. After multivariate analysis, size or generation of valves were not independent predictive factors of valve deterioration. Evolution of mean aortic gradient Conclusions After a maximal 8-year follow-up, valve deterioration after balloon-expandable TAVI is very low. In this high-risk population, TAVI seems to be a safe and durable alternative to surgery in severe aortic stenosis regardless of prosthesis generation. Acknowledgement/Funding Edwards Lifesciences