Use of a Catecholamine Sensor in the Control of an Artificial Heart System
An electrochemical sensor system to allow real-time measurement and feedback of catecholamine concentrations was developed for use in the control of artificial hearts. Electrochemical analyses were carried out using a carbon fiber working electrode, an Ag-AgCI reference electrode, and a potentiostat. The operating parameters of the pneumatically-driven artificial heart system were altered in accordance with the algorithm for changes in the catecholamine concentration. The minimum detectable concentrations of both adrenaline and noradrenaline in a mock circulatory system using a phosphate-buffered solution were approximately 1-2 ng/ml (10-8 mol/L). An artificial heart control system utilizing this set-up performed satisfactorily without delay, although sensor sensitivity decreased when placed in goat plasma instead of a phosphate-buffered solution, due to the adsorption of various substances such as plasma proteins onto the electrodes. This study demonstrated the future feasibility of a feedback control system for artificial hearts using catecholamine concentrations.