3d sequence
Recently Published Documents


TOTAL DOCUMENTS

23
(FIVE YEARS 7)

H-INDEX

7
(FIVE YEARS 0)

2021 ◽  
Vol 2021 ◽  
pp. 1-4
Author(s):  
Myriam Jrad ◽  
Haifa Zlitni ◽  
Miriam Boumediene ◽  
Atef Ben Nasr ◽  
Meriem Bouzrara

Inner ear hemorrhage is an extremely rare cause of sudden sensorineural hearing loss with few cases reported in the literature. We report the case of a 30-year-old male who presented with a sudden left ear hearing loss, with no tinnitus nor vertigo. The audiogram revealed a profound left sensorineural hearing loss. An MRI of the brain and internal auditory canal was performed 3 weeks after and revealed an increased signal intensity on T1-weighted (T1W) and T2 fluid-attenuated inversion recovery (FLAIR) images in the left cochlea. No other abnormalities were found, in particular no enhancement after intravenous administration of gadolinium. The CISS 3D sequence showed a signal of discreetly lower intensity in the left cochlea compared to the right one. The diagnosis of intracochlear hemorrhage was made. No improvement of the hearing loss has been noted after medical treatment and hyperbaric oxygen therapy.


2021 ◽  
Vol 23 (Supplement_6) ◽  
pp. vi129-vi129
Author(s):  
Lubna Hammoudeh ◽  
Ho Young Lee ◽  
Evangelia Kaza ◽  
Jeffrey Guenette ◽  
Tracy Balboni

Abstract BACKGROUND Currently, the standard MRI sequence for SC imaging in SBRT has been axial 2D T2-weighted Turbo Spin Echo (TSE). Even though 3D T2-weighted sequences such as SPACE (Sampling Perfection with Application optimized Contrasts using different flip angle Evolution) image a whole volume simultaneously and thus offer better reconstruction, they have not been clinically implemented due to their long acquisition times. However, the application of Compressed Sensing (CS) methods on SPACE sequences, achieving clinically acceptable time. METHODS A 3D T2 CS SPACE was obtained and evaluated against the standard 2D TSE for spine SBRT based on a MagPhan RT quality assurance phantom and patients data, analysis was done using the phantom manufacturer software ImageOwl that calculates image distortions by comparing the known position of phantom features to their detected position in the image. RESULTS Results of phantom comparison between 3D T2 and 2D T2 indicate that although the 3D sequence had lower signal-to-noise ratio (SNR) than the 2D sequence, it presented less geometric distortions caused by gradient non-linearities, particularly in the anterior-posterior (A/P) and head-feet (H/F) directions. Distortions caused by chemical shift are in theory smaller for the 3D T2 CS SPACE, amounting to 0.85mm compared to 1.62mm with 2D T2. Between 2D versus 3D MRI defined SC data among 4 patients, average deviation of the centroid point cord contours was 0.08cm. The volume of the cord showed 1cc larger 3D volumes compared to 2D T2. Finally, the mean voxel count overlap coefficient and DICE coefficient was 0.92 and 0.87 respectively. CONCLUSIONS Since 3D MRI is under consideration to replace 2D MRI, it is important to compare SC contours from 3D to 2D MRI and assess their impact on treatment plans. Positive results would pave the path for larger subject cohort evaluation.


Author(s):  
Christoph Berliner ◽  
Zhiyue J. Wang ◽  
Sylvia T. Singer ◽  
Regine Grosse ◽  
Rosalie V. McDonough ◽  
...  

Abstract Purpose Anterior pituitary iron overload and volume shrinkage is common in patients with transfusion-dependent anemia and associated with growth retardation and hypogonadotropic hypogonadism. We investigated the accuracy of different MRI-based pituitary volumetric approaches and the relationship between pituitary volume and MRI-R2, particularly with respect to growth and hypogonadism. Methods In 43 patients with transfusion-dependent anemia (12–38 years) and 32 healthy controls (12–72 years), anterior pituitary volume was measured by a sagittal T1 GRE 3D sequence at 1.5T and analyzed by 3D semi-automated threshold volumetry (3D-volumetry). This reference method was compared with planimetric 2D-volumetry, approximate volume calculations, and pituitary height. Using a multiple SE sequence, pituitary iron as MRI-R2 was assessed by fitting proton signal intensities to echo times. Growth and hypogonadism were obtained from height percentile tables and patients’ medical charts. From body surface area and age adjusted anterior pituitary volumes of controls, Z‑scores were calculated for all subjects. Separation of controls and patients with respect to Z and pituitary R2 was performed by bivariate linear discriminant analysis. Results Tuned 2D volumes showed highest agreement with reference 3D-volumes (bias −4.8%; 95% CI:−8.8%|−0.7%). A linear discriminant equation of Z = −17.8 + 1.45 · R2 revealed optimum threshold sensitivity and specificity of 65% and 100% for discrimination of patients from controls, respectively. Of correctly classified patients 71% and 75% showed hypogonadism and growth retardation, respectively. Conclusion Accurate assessment of anterior pituitary size requires 3D or precise 2D volumetry, with shorter analysis time for the latter. Anterior pituitary volume Z‑scores and R2 allow for the identification of patients at risk of pituitary dysfunction.


Author(s):  
Alexander H. J. Staal ◽  
Andor Veltien ◽  
Mangala Srinivas ◽  
Tom W. J. Scheenen

Abstract Purpose Isoflurane (ISO) is the most commonly used preclinical inhalation anesthetic. This is a problem in 19F MRI of fluorine contrast agents, as ISO signals cause artifacts that interfere with unambiguous image interpretation and quantification; the two most attractive properties of heteronuclear MRI. We aimed to avoid these artifacts using MRI strategies that can be applied by any pre-clinical researcher. Procedures Three strategies to avoid ISO chemical shift displacement artifacts (CSDA) in 19F MRI are described and demonstrated with measurements of 19F-containing agents in phantoms and in vivo (n = 3 for all strategies). The success of these strategies is compared to a standard Rapid Acquisition with Relaxation Enhancement (RARE) sequence, with phantom and in vivo validation. ISO artifacts can successfully be avoided by (1) shifting them outside the region of interest using a narrow signal acquisition bandwidth, (2) suppression of ISO by planning a frequency-selective suppression pulse before signal acquisition or by (3) preventing ISO excitation with a 3D sequence with a narrow excitation bandwidth. Results All three strategies result in complete ISO signal avoidance (p < 0.0001 for all methods). Using a narrow acquisition bandwidth can result in loss of signal to noise ratio and distortion of the image, and a frequency-selective suppression pulse can be incomplete when B1-inhomogeneities are present. Preventing ISO excitation with a narrow excitation pulse in a 3D sequence yields the most robust results (relative SNR 151 ± 28% compared to 2D multislice methods, p = 0.006). Conclusion We optimized three easily implementable methods to avoid ISO signal artifacts and validated their performance in phantoms and in vivo. We make recommendation on the parameters that pre-clinical studies should report in their method section to make the used approach insightful.


Author(s):  
Rodrigo Luna ◽  
Jan Fritz ◽  
Filippo del Grande ◽  
Shivani Ahlawat ◽  
Laura M. Fayad
Keyword(s):  

Author(s):  
Biccari Elia ◽  
Castellani Alessandro ◽  
Costantini Giacomo ◽  
Fiacca Andrea ◽  
Piergentini Marco

In this study we wanted to analyse how additive manufacturing could bring benefit to the treatment of intracranial aneurysms. To be able to do so we segmented Magnetic Resonance Imaging (MRI) images, specifically we used images from a Time-Of-Flight (TOF) 3D sequence. Once we obtained the virtual three-dimensional model from the diagnostic images, we proceeded to create a physical model through 3D-printing that was later used by the surgeon for a preoperative planning.


2017 ◽  
Vol 2017 ◽  
pp. 1-15
Author(s):  
Khouloud Samrouth ◽  
Olivier Deforges ◽  
Yi Liu ◽  
Mohamad Khalil ◽  
Wassim EL Falou

For some 3D applications, one may want to focus on a specific depth zone representing a region of interest in the scene. In this context, we introduce a new functionality called “autofocus” for 3D image coding, exploiting the depth map as an additional semantic information provided by the 3D sequence. The method is based on a joint “Depth of Interest” (DoI) extraction and coding scheme. First, the DoI extraction scheme consists of a precise extraction of objects located within a DoI zone, given by the viewer or deduced from an analysis process. Then, the DoI coding scheme provides a higher quality for the objects in the DoI at the expense of other depth zones. The local quality enhancement supports both higher SNR and finer resolution. The proposed scheme embeds the Locally Adaptive Resolution (LAR) codec, initially designed for 2D images. The proposed DoI scheme is developed without modifying the global coder framework, and the DoI mask is not transmitted, but it is deduced at the decoder. Results showed that our proposed joint DoI extraction and coding scheme provide a high correlation between texture objects and depth. This consistency avoids the distortion along objects contours in depth maps and those of texture images and synthesized views.


Sign in / Sign up

Export Citation Format

Share Document